General information on each IMS technology.
Abstract
The integration of ion mobility spectrometry with mass spectrometry (as an IM-MS hybrid instrument) provides additional analytical separation and identification capabilities that have greatly advanced various fields, including biomedical, pharmaceutical, and forensic sciences. In this chapter, a comprehensive exploration of various IMS instrumentation platforms is discussed, including Drift tube (DTIMS), Traveling wave (TWIMS), Trapped (TIMS), Field asymmetric waveform (FAIMS), and Differential mobility analyzer (DMA). Their respective advantages and limitations are evaluated in the context of distinct applications, including isomer separation, signal filtering to increase signal-to-noise ratio, and collision cross section (CCS) measurements in targeted and untargeted omics-based workflows. The scanning rate compatibility between various IMS devices and different mass analyzers resulted in various IM-MS hyphenation platforms. Higher sensitivity and selectivity are further achieved with the introduction of tandem IMS such as TIMS-TIMS-MS. IMS separations occur in the millisecond range and can therefore be easily incorporated into the liquid chromatography-mass spectrometry workflows and coupled with ambient ionization MS for metabolomics, lipidomic, proteomics, etc. The emergence of high-resolution IMS instruments such as Cyclic Ion Mobility Spectrometry (cIMS) and Structures for lossless ion manipulations (SLIM) is also discussed for the improvement of separation of isomers and increased predictive accuracy of CCS by machine learning models.
Keywords
- ion mobility mass spectrometry
- CCS experimental measurements
- CCS prediction models
- ambient ionization
- omics
1. Introduction
Compounds with the same mass-to-charge ratio (
2. IM-MS: Instrumentation and experimental CCS measurement
IMS separate ions based on their mobility rates (
2.1 Drift tube ion mobility spectrometry (DTIMS)
A distinctive feature of DTIMS is the uniform application of a weak electric field (10–20 V/cm) that propagates through the drift tube, which is filled with a neutral buffer gas such as nitrogen or helium that has no directional flow, as shown in Figure 1A. The two working states of DTIMS are the reduced pressure (~4 Torr e.g., Agilent 6560 IM-QTOF) and atmospheric pressure (760 Torr e.g., TOFWERK-IMS-TOF) [9]. The choice between low-pressure and high-pressure systems involves trade-offs; while low-pressure systems offer higher sensitivity and improved ion focusing due to reduced collisions frequency with the buffer gas, high-pressure IMS systems yield enhanced separation capacity due to more frequent collisions, but often experience significant ion losses [10]. Radiofrequency (RF) confinement in low-pressure DTIMS (3.5 Torr) has been reported for excellent ion transmission resulting in increased sensitivity [11]. The ability to calculate CCS from first principles using the Mason-Schamp equation is perhaps the most significant advantage of DTIMS. The utilization of a low and uniform electric field enables DTIMS to measure
![](/media/chapter/a043Y00000yJC5VQAW/a093Y00001fm3IyQAI/media/F1.png)
Figure 1.
Schematic diagram of various IMS technology and separation mechanisms: (A) DTIMS, (B) TWIMS, (C) TIMS, (D) FAIMS, and (E) DMA.
DTIMS | TWIMS | TIMS | DMA | FAIMS | |
---|---|---|---|---|---|
Separation mechanism | Temporal | Temporal | Trap & Release | Spatial | Spatial |
Gas flow to ion motion | No net flow | No net flow | Parallel | Perpendicular | Parallel |
Electric field | Static | Oscillating | Static | Static | Oscillating |
Ion injection | Pulsed ion packet | Pulsed ion packet | Variable | Continuous filter | Continuous filter |
Ion analysis | All ion analysis | All ion analysis | All ion & scannable | Scannable | Scannable |
CCS | Direct | Calibrated | Calibrated | Direct | No CCS |
Common calibrants | Agilent tune mix | Polyalanine, Waters CCS Major Mix | Agilent tune mix | ||
Resolving power | 60–80a | ~40b | ~200 | <100 | ~40c |
Vendors | Agilent, TofWerk, Excellims | Waters | Bruker | SEADM, TSI | Owlstone, Thermo, Heartland, Sciex |
Mass analyzers | TOF, Quadrupole | TOF | TOF, FTICRs, Orbitraps | TOF, FTICRs, Orbitraps | TOF, FTICRs, Orbitraps |
Table 1.
180–250 with HRdm.
cIMS and SLIM: >400.
Up to 500 with the doping of the N2 carrier gas with light gases such as He or H2.
While DTIMS systems enable comprehensive ion collection, i.e., the arrival time of all ions is recorded during each measurement cycle, particularly valuable for the analysis of complex biological samples, ions are not continuously injected. Instead, ion packets are introduced as ion pulses into the drift tube using an ion gate [15] or an ion funnel [16]. The time scale of the ion release time and IMS separation time is 25–400 μs and 25–100 ms respectively, resulting in an experimental duty cycle of ~0.04–1%. Consequently, only a fraction (0.1–1%) of the generated ions are transmitted to the drift tube for separation, resulting in a low sensitivity [17]. To improve the DTIMS sensitivity, an ion multiplexing approach is utilized, involving the pulsing of multiple packets of ions into the drift region at defined times. The multiplexing strategy results in an overlapping ion mobility spectrum, which is subsequently deconvoluted to the correct arrival times using schemes such as the Hadamard Transformation algorithm, achieving improvements as high as 50% duty cycle [5, 18]. An additional challenge of DTIMS systems is the low resolving power which can be improved using a post-acquisition data reconstruction technique or modifying the DTIMS instrumentation [5, 19]. An example of a reconstruction technique is the High-resolution demultiplexing tool (HRdm, Agilent Technologies), which effectively increases the resolving power from ~60 to a range between 180 and 250 [20]. This involves passing multiple ion packets through the drift tube, giving each ion packet a shorter accumulation time to improve the duty cycle while reducing the negative effects of space charge. The spectrum obtained is then deconvoluted and demultiplexed using the Hadamard transform algorithm, resulting in an improved signal-to-noise ratio and lower detection limits. The increase in the resolving power through instrumentation is accomplished by increasing the voltage drop across the drift cell and decreasing temperature [5]. For precise DTIMS measurements, it is essential to keep the ions in the low field limit. Therefore, to increase the voltage drop either the length of the drift cell or pressure needs to be increased [5, 21]. Examples include DTIMS systems of 2 m by Bowers et al. [22] and ~2.9 m by Clemmer et al. [23], as well as atmospheric pressure DTIMS systems yielding resolving power between 100 and 250 [9].
2.2 Traveling wave ion mobility spectrometry (TWIMS)
The drift region of TWIMS is very similar schematically to the DTIMS platform; it contains a stacked set of ring electrodes. However, in TWIMS, an RF voltage is applied across consecutive electrodes to confine the ions radially, and pulse differential current (DC) voltage is applied to push ions axially as shown in Figure 1B. Similarly to the RF-confining drift cell, the RF confinement in TWIMS focuses the ions, leading to increased analyte signals attributed to reduced ion diffusion. Under the combined actions of the applied voltages, traveling electric field waves are formed to propel the ions through the IMS tube under reduced pressure of ~2–4 Torr. The amplitude and speed of the electric waves determine the ion separation in TWIMS [14]. Due to the existence of RF voltage, ions are not in a constant field in TWIMS mode, requiring prior calibration of TWIMS instruments with ions of known mobility before determining CCS values of unknowns (analogous to the single-field CCS measurements in DTIMS) [5, 14]. Various compounds are used as calibrants in TWIMS, including polyalanine and Waters CCS Major Mix (Table 1). Since the accuracy of CCS measurement in TWIMS depends on the type of calibrants being used, class-specific calibrants are needed for more accurate CCS measurements. For example, calibrating instruments to obtain CCS values using peptides in lipid analyses has demonstrated a significant error [5, 24]. TWIMS shares both advantages and disadvantages with DTIMS, such as the full ion mobility spectrum acquisition and pulsed ion packet delivery by ion gating. Two primary advantages distinguish TWIMS: the low voltage requirements due to constant wave heights, and the ability to manipulate ion motion to long path lengths without significant ion losses. These attributes were the key to the development of two platforms with extremely long path lengths leading to very high resolving power: cIMS from Waters Corporation with a resolving power of ~750 for 100 passes [25] and SLIM, developed at PNNL and commercialized by MOBILion Systems (Chadds Ford, PA) with a path length of 13 m in a device measuring 45.9 cm × 32.5 cm, with a separating power of ~1860 [26].
Commercial TWIMS instruments are produced by Waters Corporation, which includes the Synapt G2 and Vion series (Table 1). The Synapt G2 model is designed with an IM tube positioned behind the quadrupole. The collision cell is situated both before and after the drift tube (in DTIMS the collision cell can be connected only after the IMS), enabling TWIMS-MS to perform time-aligned parallel fragmentation [3]. The Vion series moved the IM tube and its frontally connected collision cell to the front of the quadrupole. For comprehensive compound structure elucidation, in addition to the collision-induced dissociation (CID), TWIMS can combine surface-induced dissociation (SID) or electron capture dissociation (ECD), essential tools for structural details of both ligand-bound and unbound [3]. Compared to DTIMS, the resolving power of TWIMS is not significantly improved as it is less than 100.
2.3 Trapped ion mobility spectrometry (TIMS)
The separation principle of TIMS is effectively inverted from traditional ion mobility modalities. While DTIMS and TWIMS methods maintain, the essentially stationary gas flow, with an electric field parallel to the ion movement, TIMS utilizes a unidirectional buffer gas parallel to the ion motion, accompanied by an opposing electric field (ca. 70 V/cm) [3]. The TIMS analyzer is comprised of a set of electrodes that form three regions: the entrance funnel, the TIMS mobility region (ion mobility analyzer), and the exit funnel. The entrance and exit regions serve to control ion deflection and focusing, while the TIMS mobility region is utilized to accumulate, trap, and elute ions of interest which effectively increases the measurements’ duty cycles. In the TIMS mobility region, different plates along its length modulate the potential of ions as a function of distance. Consequently, ions with larger CCS values will be pushed further along the TIMS analyzer due to the energy they receive from the carrier gas as shown in Figure 1C. The accumulated ions can then be eluted from the TIMS analyzer by sequentially lowering the position-dependent plate potential as a function of time. This sequence of accumulation and elution processes can be adjusted to optimize for fast scans (tens of ms) or high-resolution separations (hundreds of ms) [5]. Optimized stepping scan functions can provide IMS resolving power of >300 while reducing overall experiment time and increasing the duty cycle [10]. An additional main difference between TIMS and the previously described DTIMS and TWIMS techniques lies in its operational scanning mode. In DTIMS and TWIMS, all ions are observed under the same experimental conditions, while TIMS requires changes to experimental parameters to elute all ions. Though TIMS instruments can indeed measure
2.4 Field asymmetric waveform ion mobility spectrometry (FAIMS)
Both DTIMS and TWIMS separate ions by the gas-phase mobility inherent in an ion for a given gas under low electric field conditions. At high electric fields (i.e., > 104 V/cm), ion mobility begins to change due to non-elastic ion-gas collisions (e.g., reactive, or interactive ion-gas collisions). An IM technique that exploits this differential mobility behavior at high fields is FAIMS operating at atmospheric pressure and near-ambient conditions [14]. Two ions with different chemical compositions can have the same ion mobility at a low electric field, but completely different mobilities at a high field, allowing them to be differentiated in the FAIMS measurement. FAIMS’s current commercial instruments include Thermo Fisher Scientific, Owlstone Medical, Heartland MS, and Sciex (Table 1). FAIMS utilizes an asymmetric waveform voltage known as Dispersion Voltage (DV) which comprises high-field components of short duration and low-field components of a longer time. As a result, ions oscillate between two electrodes in an up-and-down motion, as illustrated in Figure 1D. The dispersion voltage alone would ultimately cause all ions to contact the electrodes and be neutralized. To account for this, a second voltage – a variable compensation voltage (CV) is applied during the experiment. For ions at a specific
2.5 Differential mobility analyzer (DMA)
DMA operates similarly to DTIMS; both systems use a constant electric field and can measure CCS directly. However, DMA instruments introduce distinct features, including unidirectional gas flow (as opposed to the no-directional gas flow in DTIMS), operation at ambient pressure, and scanning for target analytes [10]. This scanning capability makes DMA a narrow bandpass IM technique, analogous in concept to FAIMS, where ions travel between two parallel electrodes in the presence of a gas flow. In DMA a constant electric field is applied between two cylindrical and concentric metal electrodes and ions are introduced at the entrance electrode where they are pushed towards the exit electrode through orthogonal sheath gas flow, as shown in Figure 1E. Only ions with the appropriate mobility will reach the cell exit, while others collide with the electrode, thus preventing their detection [8]. By scanning the electric field, an ion spectrum based on the different ion mobilities can be recorded. The well-defined nature of the electric allows for high-precision measurements of ion CCS using DMA. Notably, DMA can be utilized for analyses that are not possible with DTIMS; DMA detects very large analytes, such as antibodies, aerosol particles, viruses, and other macromolecules (~ 10’s to 100’s of nm2); and is not commonly used in screening applications for small molecules (e.g., lipids and metabolites) [5]. DMA instruments are currently marketed by SEADM and TSI (Table 1).
3. Ion mobility mass spectrometry hyphenation
3.1 Mass analyzers interfaced with IM
The difficulty of coupling an IM device to a mass analyzer is caused by the need for scanning rate compatibility of both instruments. Time of flight (TOF) mass analyzers due to their microsecond acquisition timescale are readily hyphenated with all types of IM devices, facilitating the recording of several mass spectra per mobility scan [27]. For higher MS performances in terms of dynamic range, mass resolving power, and mass accuracy, Fourier transform mass spectrometry (FTMS) analyzers (e.g., Orbitrap-based or FTICR instruments) are required. FTMS analyzers can be readily coupled with spatial dispersive IM instruments, for instance, the coupling of FAIMS with LTQ-orbitrap and FTICR instruments is routinely used [14]. Hyphenation of FTICR or orbitrap-based instruments with temporal dispersive IM systems is less straightforward due to the time scale difference between the rapid IM separation and the slow FTMS acquisition rate (up to 1 s for FTICR). To circumvent this limitation, the IM separation is “slowed down” to make it suitable for the FTMS scan rate [14]. Examples include the coupling of DTIMS with an Orbitrap-based instrument using a dual gate approach [28]. The scan rate in TIMS in theory can be conducted at any speed and synchronized with any analyzer. For instance, TIMS coupling with TOF analyzer, FTICR coupling with gated TIMS, selective accumulation-TIMS (SA-TIMS), and oversampling selective accumulation TIMS (OSA-TIMS) [14, 27]. However, TOF mass analyzers are used in all the commercial IM-MS platforms (e.g., DTIMS, TWIMS, and TIMS) due to the complexity nature of coupling with FTMS analyzers.
3.2 Front-end separations coupling with IM-MS
The incorporation of front-end separation techniques such as LC with IM-MS results in an enhanced peak capacity, detection of more features, separation of co-eluting compounds and isobaric components, and increasing S/N ratios due to the orthogonality of the separation techniques [5]. Notably, the addition of IM to LC–MS workflow has been shown to yield a 15% increase in peak capacity, as reported by Pacini et al. [29]. In the context of lipid mapping of human plasma, the implementation of 2D-LC coupled with MS (LC × LC–MS) enabled the detection of 1100 features, with 100 lipids identified. Comparatively, the LC-IM-MS approach provided 800 features, identifying 55 as lipids. Despite the fewer features detected by LC-IM-MS, it offers better structural identification and higher throughput (190- and 20-min analysis time for LC × LC–MS and LC-IM-MS, respectively) [30]. Olajide et al. utilized a multidimensional LC-IM-MS/MS method to detect features that served as strain-indicating biomarkers for efficiently discriminating
3.3 Ambient Ionization Coupling with IM-MS
Ambient ionization MS is a technique for direct analysis by ionizing analytes at ambient pressure and temperature conditions without sample preparation. The advantages of ambient ionization techniques include minimal sample preparation, low sample and solvent volume utilization, and direct rapid analysis [36]. Ambient ionization can be largely categorized into three main classes primarily based on their desorption methods: liquid extraction, plasma desorption, and laser ablation [37]. An extensive description of each ambient ionization technique under each classification and their applications in forensics, biomedical, environmental, bioanalysis, in vivo analysis, food and agriculture, and reaction monitoring and catalysis have been detailed in previous reviews [37, 38]. The direct sample analysis using ambient ionization MS suffers severe matrix interference and chemical noise due to no chromatographic separation and reduced selectivity as isomers cannot be separated [4].
The coupling of IM-MS to ambient ionization has proven to increase the technique selectivity due to isomer separation [39]. In addition, ambient ionization IM – MS is very rapid compared to classic LC – IM – MS workflows. Desorption electrospray ionization (DESI) was coupled to DTIMS to investigate charge state distributions and cross sections for protein ions in a single experiment [40]. DESI was also coupled with FAIMS to effectively increase the image quality of targeted compounds in sea algae and mouse brain tissue sections [41]. The capability of DESI to produce ions directly from tablets and creams was explored in combination with hyphenated IMS for the analysis of pharmaceutical drugs [42]. The signal filtration capability of TWIMS was demonstrated in a selective surface analysis when coupled with Desorption atmospheric pressure photoionization (DAPPI) and Direct analysis in real-time (DART) resulting in a low limit of detection (LOD) [43]. Liquid extraction surface analysis (LESA) and FAIMS were integrated to image proteins in mouse brain and liver tissue samples [44]. FAIMS has also been coupled with paper spray (PS) to obtain metabolomic and liposome characteristics in breast tissues [45]. Olajide et al. coupled PS and leaf spray (LS) to DTIMS for the separation of geometric and constitutional isomers [39]. Furthermore, the PS-IM-MS/MS developed by Olajide et al. was utilized to distinguish five
3.4 Tandem ion mobility (IMS/IMS)
The coupling of multiple ion mobility cells allows low-intensity features to be detected. Tandem ion mobility separations have been achieved either by hyphenation of the FAIMS device with conventional ion mobility geometry or by combining ion mobility cells of the same geometry [14]. Examples include FAIMS-DTIMS-MS with the FAIMS increasing the S/N ratio and the DTIMS providing accurate CCS measurement [46], DTIMS-DTIMS-MS [47], and TIMS-TIMS-MS [48]. In a study on the tryptic peptide, IMS provided a peak capacity of 60–80 while tandem IMS provided a peak capacity of 480–1360 indicating the advantage of using tandem IMS in the analysis of complex mixtures, especially in metabolomics, lipidomics, and proteomics [49]. The history of tandem IM instruments, recent developments, and detailed applications to biological systems can be found in a recent review [50].
4. High-resolution ion mobility and new technologies
The development of high-resolution IMS (HRIMS) has emerged as a solution to the low resolving power of conventional IM techniques (~60), which limits the separation of difficult isomers such as cis/trans. Achieving a resolving power of at least 300 is considered essential for effectively resolving such challenging isomers [51]. HRIMS involves advancements in technology and techniques to improve the separation and characterization of ions in complex samples. This is particularly important as samples become more complex and require more precise separation methods [52]. IMS platforms that offer high resolving power include Waters’ cyclic ion mobility spectrometry (cIMS), MOBILion Systems’ SLIM device, and the previously-described Bruker’s TIMS [10].
4.1 Cyclic ion mobility spectrometry (cIMS)
cIMS is an innovative platform that uses the TWIMS concept but has a closed-loop design with a 98 cm path length. It is comprised of an IMS separator, ion entry, and ion exit regions. This model utilizes opposite phases of RF voltage to adjacent plates in the y-axis for a pseudopotential barrier and ion confinement in the z-direction. The applied DC voltage on the “repeller” band electrodes gives ion confinement in the x-axis and TWs in a repeated pattern are supplied to the RF electrodes to propel ions forward which gets them separated as a function of their respective mobilities (Figure 2). During the injection of ions, the DC bias of the array electrodes is lower than that of the cIM electrodes, the prearray store (prior to the entrance), and the exit array regions. During IM separation, the DC offset applied to the array electrodes is increased to be similar to the cIM electrodes while being lower than that of the entrance and exit regions. Then the ions are guided through the circular path using the TW voltages applied to the cIM and array electrodes and are subsequently separated based on their ion mobility. After the separation period, the mobility-separated ions are ejected (towards the mass spectrometry) by lowering the DC offset of the array electrodes with respect to that of the cIM. However, this DC offset is still higher than the exit and postarray (after the exit). In addition, separated ions can also move to the prearray store if needed by adjusting the voltages accordingly. In short, this model provides a “multifunction” option for mobility selection, activation, storage, IMSn, as well as custom combinations of these functions.
![](/media/chapter/a043Y00000yJC5VQAW/a093Y00001fm3IyQAI/media/F2.png)
Figure 2.
cIMS instrument design. (Adapted with permission from Ref. [
Interestingly, cIMS provides a resolving power of around 80 for a single pass while 750 for 100 passes, but also loses its transmission ion efficiency with each pass (~2.4% per pass) [25]. Additionally, much work on cIMS has been conducted showing promising separation results [53, 54, 55]. For example, Williamson et al. utilized mass distribution-based isotopic shift separation for the characterization of isomeric species and conformers with very good reproducibility [56]. Harrison et al. reported the use of cIMS coupled with a temperature-controlled electrospray ionization source to analyze large biomolecular systems, temperature-induced protein aggregation, and oligonucleotide complexes [55]. Another study incorporated liquid extraction surface analysis with cIMS for the analysis of intact proteins in mouse brains and rat kidneys, increasing the number of proteins being detected with each pass using multipass experiments [54].
4.2 Structures for lossless ion manipulations (SLIM)
Nearly a decade ago, SLIM platforms were developed to demonstrate their application in high-resolution IM separation. A SLIM device consists of two parallel boards which are fabricated using printed circuit board (PCB) technology, which is fast and inexpensive. SLIM boards can also be fabricated using 3D printing or other processes. Ion motion within SLIM is driven by the action of net forces caused by the electric field (DC and/or TW) and collision with a buffer gas while efficient ion confinement is achieved by applying a combination of high frequency (RF) and DC “guard” voltages. The applied fields in SLIM create “an ion-pipe” in which the ions are transmitted “losslessly” through the device (Figure 3). Therefore, the ions can travel different paths in the SLIM by changing or removing the electric field [57]. In addition, simulations of ion motion and the applied electric fields within the SLIM with the choice of fields RF and DC are investigated using computational methods such as SIMION [58, 59].
![](/media/chapter/a043Y00000yJC5VQAW/a093Y00001fm3IyQAI/media/F3.png)
Figure 3.
The RF potentials are presented by the contour lines on the left. The central plane shows the ion confinement region (iso-potential surface forming an “ion pipe”). The right plane represents the TW potential distribution between two SLIM boards and the ion confinement region. (Adapted with permission from Ref. [
Different SLIM concepts are used for ion mobility separation based on the constant field SLIM and the traveling wave SLIM (TW-SLIM). In the constant field SLIM, the first generation of SLIM, constant DC fields (e.g. DTIMS) are used for ion motion and ion mobility separation with a resistor divider network to establish a voltage gradient across the electrodes and a capacitor chain for the superimposed Ref. [57]. However, the disadvantage of the constant field SLIM is the limitation of ion path length and field strength because longer ion path lengths (e.g., 13 m) would require extremely high voltages. In contrast, since TW-SLIM repeatedly uses the same TW sequence across the device utilizing a low voltage (e.g., 30 V) enabling the extension of the IM path length significantly without the need for high voltages. In addition, the advantages of TW-SLIM include that it does not require many electronic components (resistors and capacitors). It uses phase-shifted RF waveforms for adjacent electrodes that are in the direction of the ion path, and segmented TW electrodes that are interleaved [60, 61]. Recently, the TW-SLIM model has been modified to a multi-level TW-SLIM where multiple dual-surface SLIM boards are stacked vertically, and each level contains “ion escalators” and “ion elevators” between the SLIM levels [62]. Moreover, a resolving power of ~400 was obtained with a multilevel ion path of only 43.2 m [63]. In addition, other parameters, such as plate spacing, thickness, velocity, and amplitudes can affect the ion transfer efficiency [2, 63]. Many SLIM studies are still being studied on improvement (i.e., the charge capacity), modification of SLIM designs, and coupling with different MS platforms [64, 65, 66, 67].
5. Theoretical approaches for CCS
Even though ion mobility experiments are rapid, they still have some limitations, such as the CCS accuracy, which is affected by the calibrant mixtures available for CCS calibration and the calibration method. Another limitation is the ability to identify compounds in untargeted analysis when no reference CCS values are available for the specific chemical class of interest. To solve this problem, theoretical approaches such as computational and machine-learning prediction models have been used as alternatives. More details on different CCS prediction theoretical models were recently summarized by Kartowikromo et al. [51].
5.1 Computational models
Computational models have been used for decades to calculate CCS values. CCS Predictions using these models require conversion to 3-dimensional (3D) structures to obtain all possible conformers with consideration of variables such as bond length, and protonation/deprotonation sites [68, 69]. These conformers are then optimized to the minimum energy, which is used to calculate the CCS values on a computational software platform, such as IMoS [70, 71, 72], MOBCAL [73, 74], and Sigma [75]. Some of the recent computational platforms are Rosetta online server that includes everyone (ROSIE) [76], IMPACT [77], and colloidoscope [78]. The optimization of the conformers is performed using various theoretical approaches, including density functional theory (DFT), molecular dynamics, and molecular mechanics, while the calculation of CCS is performed using algorithms, such as projection approximation (PA), exact hard-sphere scattering (EHSS), and trajectory method (TM) [68]. One of the earliest CCS prediction methods in helium gas was based on the PA algorithm using the average projected area of a large number of orientations of the analyte [79]. Furthermore, other computational CCS calculation methods have been developed, showing some improvement compared to the previous methods, such as superposition approximation (PSA) [80], diffuse hard-sphere scattering (DHSS) [70], and colloidoscope TM [78]. Olajide et al. applied the IMoS TM method to structurally characterized two conformers of verapamil based on their calculated CCS [39]. Recently, a web application (https://rosie.graylab.jhu.edu) for CCS calculations using the projection approximation rough circular shapes (PARCS) algorithm on ROSIE was reported where the projection area of each atom with a nine points rough circle was studied. However, computational approaches usually require high CPU power (computationally intensive) to generate a whole set of compounds which can even take days even though it can accurately calculate CCS values. In addition, it is also prone to larger CCS errors, especially for flexible molecular structures [81].
5.2 Machine learning (ML) models
Machine learning-based models have the advantage of less time and CPU power for predicting multiple connections with fewer errors compared to computational methods. In fact, errors of less than 10% are obtained compared to the computational methods. These models require a training set with data obtained from experimental measurements and a validation dataset divided into an internal and an external dataset [69]. In addition, molecular descriptors (MDs) are determined for the compounds in the dataset, which are numerical values obtained from molecular structures using mathematical algorithms, such as measured values (polarity, logP, dipole moment, etc.) and theoretical values (constitution, geometry, physical chemistry, etc.). These MDs are calculated or determined using computer software or programs such as Dragon, or online databases such as Lipid Maps and the Human Metabolome Database (HMDB) [82]. The combination of different MDs can provide a “fingerprint” for the chemical property of a compound. However, these MDs do not always correctly reflect the features of a compound’s structure [83]. Moreover, the MDs are preprocessed, selected, and optimized using the preferred ML algorithm. ML algorithms can be divided into nonlinear and linear methods, such as regression models, neural networks (NN), random forest (RF), or Gradient Boosting Machine (GBM) [82]. Several ML CCS prediction models that can be used for lipidomics and metabolomics studies are publicly available and are based on support vector regression such as MetCCS [84], LipidCCS [85], AllCCS [69], CCSbase [83], etc. Bijlma et al. were the first to present an ML-based model for predicting the CCS values for small molecules based on the ANN algorithm (a type of NN algorithm) [86]. Subsequently, many neural network algorithm methods and comparisons between different neural network algorithms have been reported to be used for different omics (lipidomics, metabolomics, and proteomics), in which some methods are publicly available such as DeepCCS [87], DarkChem [88], AlphaPeptDeep [89], etc. Although ML models provide fast CCS predictions with lower errors, they still have some limitations due to the availability of databases for a variety of chemical classes, the low resolving power of IMS instruments, and the calibration methods for more accurate experimental CCS values.
6. Applications of IM-MS to lipidomics, metabolomics, and proteomics
6.1 Lipidomics
Lipids are structurally diverse with various isomers performing different biological functions which has necessitated the use of IMS for their structural elucidation and isomeric separation. IM-MS separates various lipid categories, with each category occupying a distinct
6.2 Metabolomics and proteomics
Metabolomics is a large-scale study of metabolites, that is, small molecules (<1500 Da) involved in cell function and various regulatory pathways. Biological processes often produce metabolite isomers during metabolism which common analytical methods often fail to distinguish. Hines et al. successfully determined the nitrogen CCS values of 1425 drug or drug-like metabolites using TWIMS, resolving drug metabolites protomers such as fluoroquinolone protomers in her study [101]. DTIMS was utilized in the isomeric separation of steroid metabolites [102], isomeric bile acid [103], and isobaric/isomeric biomarkers in newborn screening [104]. The Zheng-Jiang Zhu group collected more than 5000 empirical metabolites CCS values from literature to predict the CCS for more than 1.6 million small molecules thereby improving unknown metabolites annotation in untargeted metabolomics [105]. Multiplexing strategy in DTIMS has significantly improved sensitivity in metabolomics with applications such as the analysis of contaminants of emerging concerns (CECs) in human urine samples. This study introduces the first comprehensive database of DTCCSN2 values of 148 CECs, and their metabolites collected utilizing the Agilent 6560 platform [106]. PNNL Preprocessor, developed by the Smith group is used for the simplification of the multiplexed data through data interpolation, demultiplexing, multidimensional smoothing, and saturation repair functions [107]. Twenty seven diverse metabolites covering a mass range of
![](/media/chapter/a043Y00000yJC5VQAW/a093Y00001fm3IyQAI/media/F4.png)
Figure 4.
SLIM SUPER IM separation of three glycol-BA isomers. 72 m SLIM SUPER IM separation as [M + α + H + K]2+ ions (left) and 85.5 m SLIM SUPER IM separation as [M + K]+ ions (right), where M is the BA and α is the cyclodextrin. (Adapted with permission from Ref. [
Protein conformers have been studied using TIMS which was used for conformational analysis of several model peptides [115]. DTIMS was utilized to measure the gas-phase conformational populations of three proteins: ubiquitin, cytochrome c, and myoglobin reporting over 260 helium and nitrogen cross section values for the three proteins [116]. cIMS has been utilized to study the gas-phase stability of protein ions and found that the native protein conformation is stable on the order of hundreds of milliseconds [117]. The applications of DTIMS, TWIMS, FAIMS, cIMS, and TIMS for various proteomic applications have been extensively detailed in a recent review [10].
7. Conclusion
IM-MS combines structural differentiation and mass analysis, improving peak capacity, resolution, sensitivity, selectivity, and isomer separation. When coupled with chromatographic separation techniques, for instance, LC-IM-MS/MS, compound identification in targeted and untargeted omics analysis is improved using the four-dimensional data (i.e., RT, CCS,
Low resolution in IM-MS instruments is the major restriction for isomer separation. For instance, the CCS difference between stereoisomers and enantiomers is less than 1% and 0.1% respectively. IM-MS instruments like DTIMS and TWIMS have resolutions of 40–60 and can only distinguish isomers with CCS differences in the range of 1.5–3%. The development of high-resolution IM-MS instruments including SLIM and cIMS with resolving power above 400 has made it possible to separate isomers with CCS difference less than 1% and can improve the development of portable ion mobility devices, especially for on-site analysis. Experimental CCS values are needed by machine learning models to accurately predict CCS; hence the use of high-resolution IM instruments would improve the accuracy of CCS prediction since challenging isomers would be separated and their CCS would be measured accordingly. This would greatly promote the application of IM-MS in the identification of unknown compounds. Improvements in IMS resolution, standardization of instrument calibration (preferred calibrant ions, etc.), enhancement of CCS database, development of CCS processing software, and coupling with other separation and MS analysis strategies are all important factors that continuously need improvement for the advancement of IM-MS and applications in all-omics studies.
Acknowledgments
Financial support for this work was provided by funds provided by Auburn University and the NIH grant 1R35GM147225.
References
- 1.
Wu Q , Wang J-Y, Han D-Q , Yao Z-P. Recent advances in differentiation of isomers by ion mobility mass spectrometry. TrAC Trends in Analytical Chemistry. 2020; 124 :115801 - 2.
Hollerbach AL, Norheim RV, Kwantwi-Barima P, Smith RD, Ibrahim YM. A miniature multilevel structures for lossless ion manipulations ion mobility spectrometer with wide mobility range separation capabilities. Analytical Chemistry. 2021; 94 (4):2180-2188 - 3.
Liu L, Wang Z, Zhang Q , Mei Y, Li L, Liu H, et al. Ion mobility mass spectrometry for the separation and characterization of small molecules. Analytical Chemistry. 2023; 95 (1):134-151 - 4.
Olajide OE, Yi Y, Zheng J, Hamid AM. Strain-level discrimination of bacteria by liquid chromatography and paper spray ion mobility mass spectrometry. Journal of the American Society for Mass Spectrometry. 2023; 34 (6):1125-1135 - 5.
Dodds JN, Baker ES. Ion mobility spectrometry: Fundamental concepts, instrumentation, applications, and the road ahead. Journal of the American Society for Mass Spectrometry. 2019; 30 (11):2185-2195 - 6.
Rister AL, Dodds ED. Liquid chromatography-ion mobility spectrometry-mass spectrometry analysis of multiple classes of steroid hormone isomers in a mixture. Journal of Chromatography B. 2020; 1137 :121941 - 7.
Li T, Yin Y, Zhou Z, Qiu J, Liu W, Zhang X, et al. Ion mobility-based sterolomics reveals spatially and temporally distinctive sterol lipids in the mouse brain. Nature Communications. 2021; 12 (1):4343 - 8.
Kliman M, May JC, McLean JA. Lipid analysis and lipidomics by structurally selective ion mobility-mass spectrometry. Biochimica et Biophysica Acta (BBA)-Molecular and Cell Biology of Lipids. 2011; 1811 (11):935-945 - 9.
Groessl M, Graf S, Knochenmuss R. High resolution ion mobility-mass spectrometry for separation and identification of isomeric lipids. The Analyst. 2015; 140 (20):6904-6911 - 10.
Delafield DG, Lu G, Kaminsky CJ, Li L. High-end ion mobility mass spectrometry: A current review of analytical capacity in omics applications and structural investigations. TrAC Trends in Analytical Chemistry. 2022; 157 :116761 - 11.
Allen SJ, Bush MF. Radio-frequency (rf) confinement in ion mobility spectrometry: Apparent mobilities and effective temperatures. Journal of The American Society for Mass Spectrometry. 2016; 27 (12):2054-2063 - 12.
Stow SM, Causon TJ, Zheng X, Kurulugama RT, Mairinger T, May JC, et al. An interlaboratory evaluation of drift tube ion mobility–mass spectrometry collision cross section measurements. Analytical Chemistry. 2017; 89 (17):9048-9055 - 13.
Ahmed A, Cho YJ, No M-h, Koh J, Tomczyk N, Giles K, et al. Application of the Mason− Schamp equation and ion mobility mass spectrometry to identify structurally related compounds in crude oil. Analytical Chemistry. 2011; 83 (1):77-83 - 14.
Delvaux A, Rathahao-Paris E, Alves S. Different ion mobility-mass spectrometry coupling techniques to promote metabolomics. Mass Spectrometry Reviews. 2022; 41 (5):695-721 - 15.
Steiner WE, Clowers BH, Haigh PE, Hill HH. Secondary ionization of chemical warfare agent simulants: Atmospheric pressure ion mobility time-of-flight mass spectrometry. Analytical Chemistry. 2003; 75 (22):6068-6076 - 16.
Tang K, Shvartsburg AA, Lee H-N, Prior DC, Buschbach MA, Li F, et al. High-sensitivity ion mobility spectrometry/mass spectrometry using electrodynamic ion funnel interfaces. Analytical Chemistry. 2005; 77 (10):3330-3339 - 17.
Kwasnik M, Caramore J, Fernandez FM. Digitally-multiplexed nanoelectrospray ionization atmospheric pressure drift tube ion mobility spectrometry. Analytical Chemistry. 2009; 81 (4):1587-1594 - 18.
Clowers BH, Siems WF, Hill HH, Massick SM. Hadamard transform ion mobility spectrometry. Analytical Chemistry. 2006; 78 (1):44-51 - 19.
May JC, Knochenmuss R, Fjeldsted JC, McLean JA. Resolution of isomeric mixtures in ion mobility using a combined demultiplexing and peak deconvolution technique. Analytical Chemistry. 2020; 92 (14):9482-9492 - 20.
Olajide OE, Yi Y, Zheng J, Hamid AM. Species-level discrimination of microorganisms by high-resolution paper spray–Ion mobility–Mass spectrometry. International Journal of Mass Spectrometry. 2022; 478 :116871 - 21.
Jiang W, Robinson RA. Ion mobility-mass spectrometry. In: Meyers RA, editor. Encyclopedia of Analytical Chemistry: Applications, Theory and Instrumentation. New Jersey, United States of America: John Wiley & Sons, Ltd; 2013 - 22.
Kemper PR, Dupuis NF, Bowers MT. A new, higher resolution, ion mobility mass spectrometer. International Journal of Mass Spectrometry. 2009; 287 (1-3):46-57 - 23.
Merenbloom SI, Koeniger SL, Valentine SJ, Plasencia MD, Clemmer DE. IMS− IMS and IMS− IMS− IMS/MS for separating peptide and protein fragment ions. Analytical Chemistry. 2006; 78 (8):2802-2809 - 24.
Feuerstein ML, Hernández-Mesa M, Valadbeigi Y, Le Bizec B, Hann S, Dervilly G, et al. Critical evaluation of the role of external calibration strategies for IM-MS. Analytical and Bioanalytical Chemistry. 2022; 414 (25):7483-7493 - 25.
Giles K, Ujma J, Wildgoose J, Pringle S, Richardson K, Langridge D, et al. A cyclic ion mobility-mass spectrometry system. Analytical Chemistry. 2019; 91 (13):8564-8573 - 26.
Deng L, Webb IK, Garimella SV, Hamid AM, Zheng X, Norheim RV, et al. Serpentine ultralong path with extended routing (SUPER) high resolution traveling wave ion mobility-MS using structures for lossless ion manipulations. Analytical Chemistry. 2017; 89 (8):4628-4634 - 27.
Gabelica V. Ion mobility–mass spectrometry: An overview. London, United Kingdom: Royal Society of Chemistry; 2021 - 28.
Ibrahim YM, Garimella SV, Prost SA, Wojcik R, Norheim RV, Baker ES, et al. Development of an ion mobility spectrometry-orbitrap mass spectrometer platform. Analytical Chemistry. 2016; 88 (24):12152-12160 - 29.
Pacini T, Fu W, Gudmundsson S, Chiaravalle AE, Brynjolfson S, Palsson BO, et al. Multidimensional analytical approach based on UHPLC-UV-ion mobility-MS for the screening of natural pigments. Analytical Chemistry. 2015; 87 (5):2593-2599 - 30.
Baglai A, Gargano AF, Jordens J, Mengerink Y, Honing M, van der Wal S, et al. Comprehensive lipidomic analysis of human plasma using multidimensional liquid-and gas-phase separations: Two-dimensional liquid chromatography–mass spectrometry vs. liquid chromatography–trapped-ion-mobility–mass spectrometry. Journal of Chromatography. A. 2017; 1530 :90-103 - 31.
Paglia G, Stocchero M, Cacciatore S, Lai S, Angel P, Alam MT, et al. Unbiased metabolomic investigation of Alzheimer’s disease brain points to dysregulation of mitochondrial aspartate metabolism. Journal of Proteome Research. 2016; 15 (2):608-618 - 32.
Tian H, Li S, Wen H, Zhang X, Li J. Volatile organic compounds fingerprinting in faeces and urine of Alzheimer’s disease model SAMP8 mice by headspace-gas chromatography-ion mobility spectrometry and headspace-solid phase microextraction-gas chromatography-mass spectrometry. Journal of Chromatography A. 2020; 1614 :460717 - 33.
Lipok C, Hippler J, Schmitz OJ. A four dimensional separation method based on continuous heart-cutting gas chromatography with ion mobility and high resolution mass spectrometry. Journal of Chromatography A. 2018; 1536 :50-57 - 34.
Zhong X, Chen Z, Snovida S, Liu Y, Rogers JC, Li L. Capillary electrophoresis-electrospray ionization-mass spectrometry for quantitative analysis of glycans labeled with multiplex carbonyl-reactive tandem mass tags. Analytical Chemistry. 2015; 87 (13):6527-6534 - 35.
Zhang X, Romm M, Zheng X, Zink EM, Kim Y-M, Burnum-Johnson KE, et al. SPE-IMS-MS: An automated platform for sub-sixty second surveillance of endogenous metabolites and xenobiotics in biofluids. Clinical Mass Spectrometry. 2016; 2 :1-10 - 36.
Chen H, Gamez G, Zenobi R. What can we learn from ambient ionization techniques? Journal of the American Society for Mass Spectrometry. 2011; 20 (11):1947-1963 - 37.
Feider CL, Krieger A, DeHoog RJ, Eberlin LS. Ambient ionization mass spectrometry: Recent developments and applications. Analytical Chemistry. 2019; 91 (7):4266-4290 - 38.
Rankin-Turner S, Reynolds JC, Turner MA, Heaney LM. Applications of ambient ionization mass spectrometry in 2021: An annual review. Analytical Science Advances. 2022; 3 (3-4):67-89 - 39.
Olajide OE, Donkor B, Hamid AM. Systematic optimization of ambient ionization ion mobility mass spectrometry for rapid separation of isomers. Journal of the American Society for Mass Spectrometry. 2021; 33 (1):160-171 - 40.
Myung S, Wiseman JM, Valentine SJ, Takats Z, Cooks RG, Clemmer DE. Coupling desorption electrospray ionization with ion mobility/mass spectrometry for analysis of protein structure: Evidence for desorption of folded and denatured states. The Journal of Physical Chemistry B. 2006; 110 (10):5045-5051 - 41.
Bennett RV, Gamage CM, Galhena AS, Fernández FM. Contrast-enhanced differential mobility-desorption electrospray ionization-mass spectrometry imaging of biological tissues. Analytical Chemistry. 2014; 86 (8):3756-3763 - 42.
Roscioli KM, Tufariello JA, Zhang X, Li SX, Goetz GH, Cheng G, et al. Desorption electrospray ionization (DESI) with atmospheric pressure ion mobility spectrometry for drug detection. The Analyst. 2014; 139 (7):1740-1750 - 43.
Räsänen RM, Dwivedi P, Fernández FM, Kauppila TJ. Desorption atmospheric pressure photoionization and direct analysis in real time coupled with travelling wave ion mobility mass spectrometry. Rapid Communications in Mass Spectrometry. 2014; 28 (21):2325-2336 - 44.
Sarsby J, Griffiths RL, Race AM, Bunch J, Randall EC, Creese AJ, et al. Liquid extraction surface analysis mass spectrometry coupled with field asymmetric waveform ion mobility spectrometry for analysis of intact proteins from biological substrates. Analytical Chemistry. 2015; 87 (13):6794-6800 - 45.
Huang Y-C, Chung H-H, Dutkiewicz EP, Chen C-L, Hsieh H-Y, Chen B-R, et al. Predicting breast cancer by paper spray ion mobility spectrometry mass spectrometry and machine learning. Analytical Chemistry. 2019; 92 (2):1653-1657 - 46.
Zhang X, Ibrahim YM, Chen T-C, Kyle JE, Norheim RV, Monroe ME, et al. Enhancing biological analyses with three dimensional field asymmetric ion mobility, low field drift tube ion mobility and mass spectrometry (μFAIMS/IMS-MS) separations. The Analyst. 2015; 140 (20):6955-6963 - 47.
Koeniger SL, Merenbloom SI, Valentine SJ, Jarrold MF, Udseth HR, Smith RD, et al. An IMS− IMS Analogue of MS− MS. Analytical Chemistry. 2006; 78 (12):4161-4174 - 48.
Liu FC, Ridgeway ME, Park MA, Bleiholder C. Tandem trapped ion mobility spectrometry. The Analyst. 2018; 143 (10):2249-2258 - 49.
Merenbloom SI, Bohrer BC, Koeniger SL, Clemmer DE. Assessing the peak capacity of IMS− IMS separations of tryptic peptide ions in He at 300 K. Analytical Chemistry. 2007; 79 (2):515-522 - 50.
Eldrid C, Thalassinos K. Developments in tandem ion mobility mass spectrometry. Biochemical Society Transactions. 2020; 48 (6):2457-2466 - 51.
Kartowikromo KY, Olajide OE, Hamid AM. Collision cross section (CCS) measurement and prediction methods in omics. Journal of Mass Spectrometry. 2023. DOI: 10.1002/jms.4973 - 52.
Garimella SV, Ibrahim YM, Webb IK, Tolmachev AV, Zhang X, Prost SA, et al. Simulation of electric potentials and ion motion in planar electrode structures for lossless ion manipulations (SLIM). Journal of The American Society for Mass Spectrometry. 2014; 25 (11):1890-1896 - 53.
Kenderdine T, Nemati R, Baker A, Palmer M, Ujma J, FitzGibbon M, et al. High-resolution ion mobility spectrometry-mass spectrometry of isomeric/isobaric ribonucleotide variants. Journal of Mass Spectrometry: JMS. 2020; 55 (2):e4465 - 54.
Sisley EK, Ujma J, Palmer M, Giles K, Fernandez-Lima FA, Cooper HJ. LESA cyclic ion mobility mass spectrometry of intact proteins from thin tissue sections. Analytical Chemistry. 2020; 92 (9):6321-6326 - 55.
Harrison JA, Pruska A, Bittner P, Muck A, Cooper-Shepherd DA, Zenobi R. Advancing cyclic ion mobility mass spectrometry methods for studying biomolecules: Toward the conformational dynamics of mega Dalton protein aggregates. Analytical Chemistry. 2022; 94 (36):12435-12443 - 56.
Williamson DL, Nagy G. Isomer and conformer-specific mass distribution-based isotopic shifts in high-resolution cyclic ion mobility separations. Analytical Chemistry. 2022; 94 (37):12890-12898 - 57.
Ibrahim YM, Hamid AM, Deng L, Garimella SV, Webb IK, Baker ES, et al. New frontiers for mass spectrometry based upon structures for lossless ion manipulations. The Analyst. 2017; 142 (7):1010-1021 - 58.
Tolmachev AV, Webb IK, Ibrahim YM, Garimella SV, Zhang X, Anderson GA, et al. Characterization of ion dynamics in structures for lossless ion manipulations. Analytical Chemistry. 2014; 86 (18):9162-9168 - 59.
Hamid AM, Prabhakaran A, Garimella SV, Ibrahim YM, Smith RD. Characterization of applied fields for ion mobility separations in traveling wave based structures for lossless ion manipulations (SLIM). International Journal of Mass Spectrometry. 2018; 430 :8-13 - 60.
Hamid AM, Ibrahim YM, Garimella SV, Webb IK, Deng L, Chen T-C, et al. Characterization of traveling wave ion mobility separations in structures for lossless ion manipulations. Analytical Chemistry. 2015; 87 (22):11301-11308 - 61.
Hamid AM, Garimella SVB, Ibrahim YM, Deng L, Zheng X, Webb IK, et al. Achieving high resolution ion mobility separations using traveling waves in compact multiturn structures for lossless ion manipulations. Analytical Chemistry. 2016; 88 (18):8949-8956 - 62.
Ibrahim YM, Hamid AM, Cox JT, Garimella SV, Smith RD. Ion elevators and escalators in multilevel structures for lossless ion manipulations. Analytical Chemistry. 2017; 89 (3):1972-1977 - 63.
Hollerbach AL, Li A, Prabhakaran A, Nagy G, Harrilal CP, Conant CR, et al. Ultra-high-resolution ion mobility separations over extended path lengths and mobility ranges achieved using a multilevel structures for lossless ion manipulations module. Analytical Chemistry. 2020; 92 (11):7972-7979 - 64.
Huntley AP, Hollerbach AL, Prabhakaran A, Garimella SV, Giberson CM, Norheim RV, et al. Development of a structure for lossless ion manipulations (SLIM) high charge capacity array of traps. Analytical Chemistry. 2023; 95 (9):4446-4453 - 65.
Kinlein ZR, Anderson GA, Clowers BH. Accelerating prototyping experiments for traveling wave structures for lossless ion manipulations. Talanta. 2022; 244 :123446 - 66.
Eaton RM, Zercher BP, Wageman A, Bush MF. A flexible, modular platform for multidimensional ion mobility of native-like ions. Journal of the American Society for Mass Spectrometry. 2023; 34 (6):1175-1185 - 67.
Hollerbach AL, Conant CR, Nagy G, Monroe ME, Gupta K, Donor M, et al. Dynamic time-warping correction for shifts in ultrahigh resolving power ion mobility spectrometry and structures for lossless ion manipulations. Journal of the American Society for Mass Spectrometry. 2021; 32 (4):996-1007 - 68.
Laphorn C, Pullen F, Chowdhry B. Ion mobility spectrometry mass spectrometry of small molecules: Separating and assigning structures to ions. Mass Spectrometry Reviews. 2013; 32 (1):43-71 - 69.
Zhou Z, Tu J, Zhu Z-J. Advancing the large-scale CCS database for metabolomics and lipidomics at the machine-learning era. Current Opinion in Chemical Biology. 2018; 42 :34-41 - 70.
Larriba C, Hogan CJ Jr. Ion mobilities in diatomic gases: Measurement versus prediction with non-specular scattering models. The Journal of Physical Chemistry A. 2013; 117 (19):3887-3901 - 71.
Larriba C, Hogan CJ Jr. Free molecular collision cross section calculation methods for nanoparticles and complex ions with energy accommodation. Journal of Computational Physics. 2013; 251 :344-363 - 72.
Larriba-Andaluz C, Fernandez-Garcia J, Ewing MA, Hogan CJ, Clemmer DE. Gas molecule scattering & ion mobility measurements for organic macro-ions in He versus N 2 environments. Physical Chemistry Chemical Physics. 2015; 17 (22):15019-15029 - 73.
Mesleh MF, Hunter J, Shvartsburg A, Schatz GC, Jarrold M. Structural information from ion mobility measurements: Effects of the long-range potential. The Journal of Physical Chemistry. 1996; 100 (40):16082-16086 - 74.
Shvartsburg AA, Jarrold MF. An exact hard-spheres scattering model for the mobilities of polyatomic ions. Chemical Physics Letters. 1996; 261 (1-2):86-91 - 75.
Wyttenbach T, von Helden G, Batka JJ Jr, Carlat D, Bowers MT. Effect of the long-range potential on ion mobility measurements. Journal of the American Society for Mass Spectrometry. 1997; 8 (3):275-282 - 76.
Turzo SB, Seffernick J, Lyskov S, Lindert S. Predicting ion mobility collision cross sections using projection approximation with ROSIE-PARCS webserver. Briefings in Bioinformatics. Oxford, United Kingdom: Oxford University Press; 2023. DOI: 10.1093/bib/bbad308 - 77.
Paizs B. A divide-and-conquer approach to compute collision cross sections in the projection approximation method. International Journal of Mass Spectrometry. 2015; 378 :360-363 - 78.
Ewing SA, Donor MT, Wilson JW, Prell JS. Collidoscope: An improved tool for computing collisional cross-sections with the trajectory method. Journal of The American Society for Mass Spectrometry. 2017; 28 (4):587-596 - 79.
von Helden G, Hsu MT, Gotts N, Bowers MT. Carbon cluster cations with up to 84 atoms: Structures, formation mechanism, and reactivity. The Journal of Physical Chemistry. 1993; 97 (31):8182-8192 - 80.
Bleiholder C, Wyttenbach T, Bowers MT. A novel projection approximation algorithm for the fast and accurate computation of molecular collision cross sections (I) method. International Journal of Mass Spectrometry. 2011; 308 (1):1-10 - 81.
Paglia G, Kliman M, Claude E, Geromanos S, Astarita G. Applications of ion-mobility mass spectrometry for lipid analysis. Analytical and Bioanalytical Chemistry. 2015; 407 :4995-5007 - 82.
Li X, Wang H, Jiang M, Ding M, Xu X, Xu B, et al. Collision cross section prediction based on machine learning. Molecules. 2023; 28 (10):4050 - 83.
Ross DH, Cho JH, Xu L. Breaking down structural diversity for comprehensive prediction of ion-neutral collision cross sections. Analytical Chemistry. 2020; 92 (6):4548-4557 - 84.
Zhou Z, Xiong X, Zhu Z-J. MetCCS predictor: A web server for predicting collision cross-section values of metabolites in ion mobility-mass spectrometry based metabolomics. Bioinformatics. 2017; 33 (14):2235-2237 - 85.
Zhou Z, Tu J, Xiong X, Shen X, Zhu Z-J. LipidCCS: Prediction of collision cross-section values for lipids with high precision to support ion mobility–mass spectrometry-based lipidomics. Analytical Chemistry. 2017; 89 (17):9559-9566 - 86.
Bijlsma L, Bade R, Celma A, Mullin L, Cleland G, Stead S, et al. Prediction of collision cross-section values for small molecules: Application to pesticide residue analysis. Analytical Chemistry. 2017; 89 (12):6583-6589 - 87.
Plante P-L, Francovic-Fontaine É, May JC, McLean JA, Baker ES, Laviolette F, et al. Predicting ion mobility collision cross-sections using a deep neural network: DeepCCS. Analytical Chemistry. 2019; 91 (8):5191-5199 - 88.
Colby SM, Nuñez JR, Hodas NO, Corley CD, Renslow RR. Deep learning to generate in silico chemical property libraries and candidate molecules for small molecule identification in complex samples. Analytical Chemistry. 2019; 92 (2):1720-1729 - 89.
Zeng W-F, Zhou X-X, Willems S, Ammar C, Wahle M, Bludau I, et al. AlphaPeptDeep: A modular deep learning framework to predict peptide properties for proteomics. Nature Communications. 2022; 13 (1):7238 - 90.
Tu J, Zhou Z, Li T, Zhu Z-J. The emerging role of ion mobility-mass spectrometry in lipidomics to facilitate lipid separation and identification. TrAC Trends in Analytical Chemistry. 2019; 116 :332-339 - 91.
Kyle JE, Zhang X, Weitz KK, Monroe ME, Ibrahim YM, Moore RJ, et al. Uncovering biologically significant lipid isomers with liquid chromatography, ion mobility spectrometry and mass spectrometry. The Analyst. 2016; 141 (5):1649-1659 - 92.
Dong L, Shion H, Davis RG, Terry-Penak B, Castro-Perez J, van Breemen RB. Collision cross-section determination and tandem mass spectrometric analysis of isomeric carotenoids using electrospray ion mobility time-of-flight mass spectrometry. Analytical Chemistry. 2010; 82 :9014-9021, 9021 - 93.
Wojcik R, Webb IK, Deng L, Garimella SV, Prost SA, Ibrahim YM, et al. Lipid and glycolipid isomer analyses using ultra-high resolution ion mobility spectrometry separations. International Journal of Molecular Sciences. 2017; 18 (1):183 - 94.
Jeanne Dit Fouque K, Ramirez CE, Lewis RL, Koelmel JP, Garrett TJ, Yost RA, et al. Effective liquid chromatography–trapped ion mobility spectrometry–mass spectrometry separation of isomeric lipid species. Analytical Chemistry. 2019; 91 (8):5021-5027 - 95.
Bowman AP, Abzalimov RR, Shvartsburg AA. Broad separation of isomeric lipids by high-resolution differential ion mobility spectrometry with tandem mass spectrometry. Journal of the American Society for Mass Spectrometry. 2017; 28 (8):1552-1561 - 96.
May JC, Goodwin CR, Lareau NM, Leaptrot KL, Morris CB, Kurulugama RT, et al. Conformational ordering of biomolecules in the gas phase: Nitrogen collision cross sections measured on a prototype high resolution drift tube ion mobility-mass spectrometer. Analytical Chemistry. 2014; 86 (4):2107-2116 - 97.
Picache JA, Rose BS, Balinski A, Leaptrot KL, Sherrod SD, May JC, et al. Collision cross section compendium to annotate and predict multi-omic compound identities. Chemical Science. 2019; 10 (4):983-993 - 98.
Paglia G, Williams JP, Menikarachchi L, Thompson JW, Tyldesley-Worster R, Halldórsson S, et al. Ion mobility derived collision cross sections to support metabolomics applications. Analytical Chemistry. 2014; 86 (8):3985-3993 - 99.
Hines KM, Herron J, Xu L. Assessment of altered lipid homeostasis by HILIC-ion mobility-mass spectrometry-based lipidomics. Journal of Lipid Research. 2017; 58 (4):809-819 - 100.
Hernández-Mesa M, Le Bizec B, Monteau F, García-Campaña AM, Dervilly-Pinel G. Collision cross section (CCS) database: An additional measure to characterize steroids. Analytical Chemistry. 2018; 90 (7):4616-4625 - 101.
Hines KM, Ross DH, Davidson KL, Bush MF, Xu L. Large-scale structural characterization of drug and drug-like compounds by high-throughput ion mobility-mass spectrometry. Analytical Chemistry. 2017; 89 (17):9023-9030 - 102.
Davis DE Jr, Leaptrot KL, Koomen DC, May JC, Cavalcanti GA, Padilha MC, et al. Multidimensional separations of intact phase II steroid metabolites utilizing LC–ion mobility–HRMS. Analytical Chemistry. 2021; 93 (31):10990-10998 - 103.
Zheng X, Smith FB, Aly NA, Cai J, Smith RD, Patterson AD, et al. Evaluating the structural complexity of isomeric bile acids with ion mobility spectrometry. Analytical and Bioanalytical Chemistry. 2019; 411 :4673-4682 - 104.
Dodds JN, Baker ES. Improving the speed and selectivity of newborn screening using ion mobility spectrometry–mass spectrometry. Analytical Chemistry. 2021; 93 (51):17094-17102 - 105.
Zhou Z, Luo M, Chen X, Yin Y, Xiong X, Wang R, et al. Ion mobility collision cross-section atlas for known and unknown metabolite annotation in untargeted metabolomics. Nature Communications. 2020; 11 (1):4334 - 106.
Belova L, Caballero-Casero N, van Nuijs AL, Covaci A. Ion mobility-high-resolution mass spectrometry (IM-HRMS) for the analysis of contaminants of emerging concern (CECs): Database compilation and application to urine samples. Analytical Chemistry. 2021; 93 (16):6428-6436 - 107.
Bilbao A, Gibbons BC, Stow SM, Kyle JE, Bloodsworth KJ, Payne SH, et al. A preprocessing tool for enhanced ion mobility–mass spectrometry-based omics workflows. Journal of Proteome Research. 2021; 21 (3):798-807 - 108.
Causon TJ, Si-Hung L, Newton K, Kurulugama RT, Fjeldsted J, Hann S. Fundamental study of ion trapping and multiplexing using drift tube-ion mobility time-of-flight mass spectrometry for non-targeted metabolomics. Analytical and Bioanalytical Chemistry. 2019; 411 :6265-6274 - 109.
Adams KJ, Ramirez CE, Smith NF, Muñoz-Muñoz AC, Andrade L, Fernandez-Lima F. Analysis of isomeric opioids in urine using LC-TIMS-TOF MS. Talanta. 2018; 183 :177-183 - 110.
Zhang W, Lin L, Xia L, Cai W, Dai W, Zou C, et al. Multi-omics analyses of human colorectal cancer revealed three mitochondrial genes potentially associated with poor outcomes of patients. Journal of Translational Medicine. 2021; 19 (1):273 - 111.
Cooper-Shepherd DA, Olivos HJ, Wu Z, Palmer ME. Exploiting self-association to evaluate enantiomeric composition by cyclic ion mobility–mass spectrometry. Analytical Chemistry. 2022; 94 (23):8441-8448 - 112.
McCullagh M, Giles K, Richardson K, Stead S, Palmer M. Investigations into the performance of travelling wave enabled conventional and cyclic ion mobility systems to characterise protomers of fluoroquinolone antibiotic residues. Rapid Communications in Mass Spectrometry. 2019; 33 :11-21 - 113.
Higton D, Palmer ME, Vissers JP, Mullin LG, Plumb RS, Wilson ID. Use of cyclic ion mobility spectrometry (cIM)-mass spectrometry to study the intramolecular transacylation of diclofenac acyl glucuronide. Analytical Chemistry. 2021; 93 (20):7413-7421 - 114.
Chouinard CD, Nagy G, Webb IK, Garimella SVB, Baker ES, Ibrahim YM, et al. Rapid Ion mobility separations of bile acid isomers using cyclodextrin adducts and structures for lossless ion manipulations. Analytical Chemistry. 2018; 90 (18):11086-11091 - 115.
Silveira JA, Ridgeway ME, Park MA. High resolution trapped ion mobility spectrometery of peptides. Analytical Chemistry. 2014; 86 (12):5624-5627 - 116.
May JC, Jurneczko E, Stow SM, Kratochvil I, Kalkhof S, McLean JA. Conformational landscapes of ubiquitin, cytochrome c, and myoglobin: Uniform field ion mobility measurements in helium and nitrogen drift gas. International Journal of Mass Spectrometry. 2018; 427 :79-90 - 117.
Eldrid C, Ujma J, Kalfas S, Tomczyk N, Giles K, Morris M, et al. Gas phase stability of protein ions in a cyclic ion mobility spectrometry traveling wave device. Analytical Chemistry. 2019; 91 (12):7554-7561